Article Text

Download PDFPDF
Influence of skin colour on diagnostic accuracy of the jaundice meter JM 103 in newborns
  1. S Samiee-Zafarghandy1,2,
  2. J Feberova1,
  3. K Williams3,
  4. A S Yasseen3,
  5. S L Perkins4,
  6. B Lemyre1
  1. 1Department of Obstetrics, Gynecology and Newborn Care, The Ottawa Hospital, Ottawa, Canada
  2. 2Department of Pediatrics, Division of Neonatology, McMaster Children's Hospital, Hamilton, Canada
  3. 3Clinical Research Unit, Children's Hospital of Eastern Ontario, Ottawa, Canada
  4. 4Department of Pathology and Laboratory Medicine, The Ottawa Hospital, Ottawa, Canada
  1. Correspondence to Dr Brigitte Lemyre, Department of Obstetrics, Gynecology and Newborn Care, The Ottawa Hospital, 501 Smyth Road, Ottawa, Canada K1H 8L6; blemyre{at}toh.on.ca

Abstract

Aim To assess the diagnostic accuracy of the JM 103 as a screening tool for neonatal jaundice and explore differential effects based on skin colour.

Methods We prospectively compared the transcutaneous bilirubin (TcB) and serum bilirubin (TSB) measurements of newborns over a 3 month-period. Skin colour was assigned via reference colour swatches. Diagnostic measures of the TcB/TSB comparison were made and clinically relevant TcB cut-off values were determined for each skin colour group.

Results 451 infants (51 light, 326 medium and 74 dark skin colour) were recruited. The association between TcB and TSB was high for all skin colours (rs>0.9). The Bland-Altman analysis showed an absolute mean difference between the two measures of 13.3±26.4 µmol/L with broad limits of agreement (−39.4–66.0 µmol/L), with TcB underestimating TSB in light and medium skin colours and overestimating in dark skin colour. Diagnostic measures were also consistently high across skin colours, with no clinically significant differences observed.

Conclusions The JM 103 is a useful screening tool to identify infants in need of serum bilirubin, regardless of skin colour. The effect of skin colour on the accuracy of this device at high levels of serum bilirubin could not be assessed fully due to small numbers in the light and dark groups.

  • neonatal jaundice
  • transcutaneous bilirubinometer

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Linked Articles

  • Fantoms
    Ben Stenson